ND 1300 QUADRA-CHEK

    Spécifications techniques

    ND 1302

    ND 1303

    ND 1304

    Axes*

    2 (XY)

    3 (XYQ) or 3 (XYZ)

    4 (XYZQ)

    Encoder inputs*

    Input frequency

    1 VPP or TTL (other interfaces upon request)

    1 VPP: 275 kHz; TTL: 3 MHz

    Subdivision factor

    10-fold (only at 1 VPP)

    Display step1)

    Adjustable, max. 7 digits

    Linear axes XYZ: 1 mm to 0.000 1 mm; angular axis Q: 1° to 0.000 1° (00° 00’ 01”)

    Display

    8.4” color flat-panel display (touchscreen); resolution: SVGA 800 x 600 pixels, for position values, dialogs and inputs, graphics functions, soft keys, and display of video images with the Video option

    Functions

    • Measurement of two-dimensional features (2-D)
    • Point measurement with crosshairs
    • Programming of features and parts
    • Measure Magic: automatic recognition of geometries
    • Tolerance display and graphic display of measurement results

    Edge finder* (option)

    • Automatic point measurement via optical edge finder

    Video* (option)

    • Automatic point measurement via video edge detection
    • Show live images
    • Archive and output live images (Archive option, only with the Video and Zoom options)
    • Zoom and light control, programmable (Zoom option, only with the Video and CNC options)

    CNC* (option)

    • Automation of measurement tasks
    • Axis control (for XYZQ) for servo and stepper motors
    • Autofocus via stepper-motor control

    Error compensation

    • Linear, and segmented linear over up to 1000 points
    • Squareness calibration; matrix compensation over up to 30 x 30 points

    Data interface

    RS-232-C/V.24 and USB (type A)

    Other connections

    • Foot switch for two functions
    • Video connection for S-Video and composite (Video option)
    • Light control over six light sources and zoom control (for Video and Zoom options)
    • CNC outputs and inputs for joystick (for CNC option)

    Accessories

    Foot switch, fiber-optic cable, holder, amplifier, calibration standard, demo parts, protective cover. tilting base, mounting adapter

    Power connection

    AC 100 V to 240 V (–15 % to +10 %), 47 Hz to 63 Hz, 100 W

    Operating temperature

    0 °C to 45 °C; (storage temperature –20 °C to 70 °C)

    Protection EN 60 529

    IP 00, front panel IP 40

    Mounting*

    Tilting base or mounting base

    Mass

    ND with tilting base:  4.8 kg; ND with mounting adapter:  2 kg

    * Please select when ordering; the Edge finder and Video options cannot be combined

    1) Depends on the signal period of the connected encoder as well as the subdivision factor